NI TB-2708 Installation Guide

6

ni.com

Measuring Floating Signals

You can use the TB-2708 to measure floating and ground-referenced
analog input signals. To measure floating signal sources, move the switch
located under the rubber-sealed door on the enclosure cover for the AI
channel you are using to the ON (floating source) switch position. In the
floating source switch position, the amplifier negative terminal connects to
ground through a 5 k

Ω resistor in parallel with a 0.1 μF capacitor. Table 2

shows the TB-2708 switch configuration options.

Measuring Ground-Referenced Signals

It is possible to set the switch on the TB-2708 to either the floating or
ground-referenced source position to measure ground-referenced signals.
However, for best results, use the OFF (ground-referenced source) switch
position to avoid ground loops. Refer to the S Series User Manual for more
information on measuring floating and ground-referenced signals.

Table 2. Configuration Summary

TB-2708 Switch

Configuration

Signal Source Types for S Series Devices

Floating Source

Ground-Referenced Source

Floating Source

Ground-
Referenced Source

+

Signal

Source

TB-2708

DAQ Device

(Differential Input

Mode)

AI GND

+

Recommended

+

Signal

Source

TB-2708

DAQ Device

(Differential Input

Mode)

AI GND

+

Not Recommended

+

Signal

Source

TB-2708

Improper Configuration

DAQ Device

(Differential Input

Mode)

+

AI GND

+

Signal

Source

TB-2708

DAQ Device

(Differential Input

Mode)

AI GND

+

Recommended