National Instruments NI 9229 User Manual

Page 16

Advertising
background image

NI 9229/9239

16

ni.com

the modules must share a single master timebase source. Refer to
the software help for information about configuring the master
timebase source for the NI 9229/9239. Visit

ni.com/info

and

enter

cseriesdoc

for information about C Series documentation.

The following equation provides the available data rates of the
NI 9229/9239:

f

s

=

where n is any integer from 1 to 31.

However, the data rate must remain within the appropriate data rate
range. Refer to the

Specifications

section for more information

about the data rate range. When using the internal master timebase
of 12.8 MHz, the result is data rates of 50 kS/s, 25 kS/s, 16.67 kS/s,
and so on down to 1.613 kS/s, depending on the value of n. When
using an external timebase with a frequency other than 12.8 MHz,
the NI 9229/9239 has a different set of data rates.

Note

The cRIO-9151 R Series Expansion chassis does

not support sharing timebases between modules.

f

M

256

÷

n

---------------------

Advertising
This manual is related to the following products: