Safety Voltages

Connect only voltages that are within the following limits.

Channel-to-earth ground isolation

Continuous ................................. 250 V

rms

,

Measurement Category II

Withstand.................................... 2,300 V

rms

, verified by a 5 s

dielectric withstand test

Channel-to-channel isolation

Continuous ................................. 250 V

rms

,

Measurement Category II

Withstand.................................... 1,390 V

rms

, verified by a 5 s

dielectric withstand test

Zone 2 hazardous locations applications in Europe

Maximum voltage
(channel-to-earth ground
and channel-to-channel) ............. 60 VDC,

Measurement Category I

Advertising