National Instruments NI 9229 User Manual

Page 21

Advertising
background image

© National Instruments Corp.

21

NI 9229/9239

Accuracy, NI 9239

Input noise

NI 9229....................................... 320

μV

rms

NI 9239....................................... 70

μV

rms

Stability

Gain drift .................................... ±5 ppm/°C
Offset drift

NI 9229 ................................ ±150

μ

V/°C

NI 9239 ................................ ±26

μ

V/°C

Measurement Conditions

Percent of

Reading

(Gain Error)

Percent of

Range

*

(Offset Error)

Calibrated max (–40 to 70 °C)

±0.13%

±0.06%

Calibrated typ (25 °C, ±5 °C)

±0.03%

±0.008%

Uncalibrated max (–40 to 70 °C)

±1.4%

±0.70%

Uncalibrated typ (25 °C, ±5 °C)

±0.3%

±0.11%

*

Range equals 10.52 V

Advertising
This manual is related to the following products: