Specifications

CLKIN Characteristics

CLKIN fundamental
frequency range

1

.................................... 1 MHz to 105 MHz,

sine or square wave

Input impedance ..................................... 50

Ω, nominal

Input coupling ........................................ AC

Voltage range

DC ................................................... ±20 V

AC ................................................... 400 mV

p-p

to 5 V

p-p

Absolute maximum input voltage

2

......... ±26 V, max

CLKIN to PXI_CLK10_IN delay
without PLL ........................................... 14 ns to 14.7 ns, typical

CLKIN to PXI_CLK10 delay
with PLL ................................................ ±1 ns, max

CLKIN frequency accuracy requirement

For PLL and TCXO ........................ ±5.0 ppm

For replacing PXI_CLK10
(no PLL).......................................... ±100 ppm

3

1

CLKIN fundamental frequency can be any multiple of 1 MHz within the range specified when the PLL is engaged and
PXI_CLK10 is locking to it. The frequency must be 10 MHz when replacing PXI_CLK10 without the PLL.

2

Stresses beyond those listed can cause permanent damage to the device. Exposure to absolute maximum rated conditions for

extended periods of time can affect device reliability. Functional operation of the device outside the conditions indicated in
the operational parts of the specification is not implied.

3

This is a requirement of the PXI specification.