9.9

Automatic Emissivity Setting

This function is used to make an automatic setting of emissivity using an external thermocouple connected with the
auxiliary input.

If necessary, enable the thermocouple input (see par. 9.2) and select the thermocouple type (see par. 8.6)

Press the <▲> or <▼> key until the thermocouple temperature measure is displayed

e

°C

Lock

Press

<ENTER> key for setting the emissivity value with reference to the temperature values read from thermocouple

and pyrometer. The emissivity will change automatically to the correct value by the OSP.

e

°C

Lock

An error message appear if the OSP measure a temperature highest of the thermocouple. If the error happen, verify

the contact between the thermocouple and the object.