9 firmware revision number, 10 instrument serial number – Omega OSP SERIES User Manual

Page 30

Advertising
background image

30

e

Set

°C

e

Set

°C

Press the <▲> or <▼> key to enable (or to disable) the thermocouple auxiliary input operation.

e

Set

°C

With the compensation function activated, press <ENTER> to begin the measurements in order to calculate the

ambient temperature average value. Point OSP towards objects and surfaces surrounding the target, making a scan
over those items and acquiring corresponding temperature values. A counter (from 1 to 255) indicates the number of
measurements considered (

OSP

displays the first value read during this procedure). These acquisitions have to be

done with emissivity

ε

=1, hence when acquisition starts the emissivity value has set automatically to one.

Press

<ENTER> to stop acquisition and to confirm the read ambient temperature (at this moment

OSP

will display

the last acquired temperature value).

Press the <SEL> key to select the new configuration step or press <Trigger> to switch the instrument off. The new

selections will be automatically memory stored.

To execute the temperature measurements with TAM, switch

OSP

on and set the emissivity to the correct target value.

10.9

Firmware revision number

Press the <SEL> key until the following page is displayed :

e

Set

°C

Lock

The above page indicates the version of the firmware installed in your instrument.
This information is extremely important in all inquires for technical assistance support.

10.10

Instrument serial number

Press the <SEL> key until the following page is displayed :

e

Set

°C

Lock

The above page indicates the serial number of the instrument. Such number is extremely important for any request of
technical assistance support.