6 buzzer on/off, 8 ambient temperature compensation (tam) – Omega OSP SERIES User Manual

Page 29

Advertising
background image

29

10.6

Buzzer On/Off

If the Alarms option is installed, it is possible to switch the buzzer on or off. This page allows the set-up of the acoustic
signal.

e

Set

°C

Lock

e

Set

°C

Lock

When the “Buzzer” is switched -Off- the alarm conditions will be announced only with the relevant -HAL- or -LAL-
message on the display.

10.7

OSP 500/800 decimal point selection

N

OTE

:

THIS FEATURE WORKS ON

OSP 500

OR

OSP800

WITH FIRMWARE FROM VERSION

4.001.

It is possible to display the temperature measured with a decimal digit.
The following procedure is used to enable the operating mode.

"

Switch the instrument off.

"

Keep

pressed

<*>+<SEL> keys and switch the unit on by pressing the <Trigger> key. The temperature

measurements will be displayed with the decimal digit.

Repeat the above procedure to disable the operating mode.

10.8

Ambient Temperature Compensation (TAM)

This function is used to compensate the measure from the ambient temperature.

Targets which have low emissivity will reflect energy from their surrounding environments, which is going to add to their
own emitted energy. If the temperature of the surrounding environment is the same value as the internal temperature of
the unit, there is no need to consider this value as it is automatically and continually measured. However, in some
industrial situations the surrounding environment (machines, furnaces, or other heat sources) has a much higher value
than the one of the internal of the unit temperature. The table below "Effect of TAM on Accuracy" underlines these effects
on measurement accuracy, with and without the use of the Ambient Temperature Compensation mode.

The Ambient Temperature Compensation mode should be used whenever the stated following points are true :
1. The target has a low emissivity.

2. Temperatures of objects facing targets are much hotter than the temperature of the unit.

3. An increased measurement accuracy is required

Actual Target

Temp.

Emissivity

Ambient Temp.

Indicated Temp.

w/o TAM

Indicated Temp.

w/ TAM

°C

°F

°C

°F

°C

°F

°C

°F

100

200

0.8

25

80

100

200

100

200

100

200

0.8

40

100

102

203

100

200

100

200

0.8

150

300

120

250

100

200

100

200

0.6

150

300

160

321

100

200

Note :

The internal temperature of IR thermometer is assumed to be 25 °C (80 °F)

Press the <SEL> key until one of the following indications is displayed :