Sect

ion 4

APPLICATION SETTING EX

AMPLE

S

Section 4

Measuring the Reference Difference between Steps

4-9

ZS-MDC

User’s Manual

[SENSING]-[CALC]-[CALC]-[STEP]

3.

Set the expression of TASK1.

INPUT X: INPUT A
INPUT Y: INPUT B

4.

Following the same procedure as 2 to 3, set
up TASK2.

INPUT X: INPUT A
INPUT Y: INPUT C

• To output the judgment result of the operation

[I/O SET]-[I/O SET]-[OUT]

5.

Select TASK1 or TASK2 depending on the
details to be output.

To output judgment on the difference between point A (reference point) and point
B: Set TASK1 to I/O SET
To output judgment on the difference between point A (reference point) and point
C: Set TASK2 to I/O SET

• To linear-output the operation result

[I/OSET]-[I/O SET]-[ANALOG]

5.

Select TASK1 or TASK2 depending on the
details to be output.

To output the difference between point A (reference point) and point B: Set TASK1
to I/O SET
To output the difference between point A (reference point) and point C: Set TASK2
to I/O SET

The calculation results of each CH can be batch-acquired if you use the communication command.

1

INPUT X

2

INPUT Y

Input A

Input B

Input C

Y X

Input A

Input B

Input C

X

Y

1

TASK1

3

TASK3

2

TASK2

4

TASK4

1

TASK1

3

TASK3

2

TASK2

4

TASK4