To run the 1a-08 hdcp compliance test using the, Srm disc – Quantum Data 882E User Manual

Page 14

Advertising
background image

13

882E Quick Start Guide - HDCP Compliance Testing with Encryption Status Tester (EST) - P/N 68-00210 Rev. B

To run the 1A-08 HDCP compliance test using the SRM disc:

1. Set up the test as described in

To run the 1A and 1B series HDCP compliance tests

on an HDMI source (transmitter):

” on page 7 using the instructions for the 882E

analyzer.

Note: The SRM test is only supported on the 882E analyzer. You can insert the SRM
disc into the source device and run the HDCP tests in batch mode rather than through
the command line.

2. Insert the SRM disc into the optical drive.

Note: The SRM test does not apply for source devices that cannot play a DVD.

3. To run 1A-08 test, use the command line as shown below:

CPTX:DUTT 1

// Selects the device type as a source.

CPTX:CPTR 8

// Selects specific test (e.g. 1A_08).

CPTX:CPTU

// Initiates the execution of the test.

Refer to

To view the HDCP compliance report:

” on page 42 for procedures on how to

view the generated report.

14

1B_05 (Transmitter Downstream
w/Repeater)

Irregular Procedure: (Second Part of
Authentication) MAX_DEVS EXCEEDED

15

1B_06 (Transmitter Downstream
w/Repeater)

Irregular Procedure: (Second Part of
Authentication) MAX_CASCADE EXCEEDED