Microlife AG1-10 User Manual

Page 14

Advertising
background image

11

improper handling, accidents, not following the operating instructions or alterations made to the

instrument by third parties. The guarantee is only valid upon presentation of the guarantee card

filled out by the dealer.

Name and company address of the responsible dealer:

8. Reference to standards

Device standard:

Device corresponds to the requirements of the

EN1060-1 /-2; ANSI / AAMI SP09

This device complies with the requirements of the Medical Device Directive 93/42/EEC.

9. www.microlife.com

Detailed user information about our products as well as services can be found at www.microlife.com

10. Technical specifications

Weight:

360 g

Size:

175 x 70 x 103 mm

Storage temperature:

­20 °C to +70 °C; 85 % relative humidity maximum

Operation temperature: 0 °C to 46 °C

Measuring range:

0 to 300 mmHg

Measuring resolution: 2 mmHg

Accuracy:

within ± 3 mmHg in 18 °C to 33 °C;

within ± 6 mmHg in 34 °C to 46 °C

Inflation source:

a volume of at least 200cc to a pressure of 300 mmHg in 4 to 10 sec

Air leakage:

<± 4 mmHg/min

Hysteresis error:

within 0 mmHg to 4 mmHg

Accessories:

1. cuff (with arm circumference of 25 - 40 cm) with inlaid bladder

2. bulb and valve

3. soft bag

Technical alterations reserved.

Advertising