6 result, Auto free switch-off (cont), Auto free – Ohaus MB45 MOISTURE ANALYZER Manual User Manual
Page 38

36
MB45
Auto free switch-off (cont)
Using the arrow buttons, scroll to MG, now enter the weight loss (1 mg
to 10 mg).
AUTO FREE
SWITCH OFF AT LESS THAN
5 MG IN
60
SECONDS
EXIT WITHOUT SAVING
Using the arrow buttons, enter the time (10 seconds to 120 seconds).
Press the Enter button.
6.6 Result
Displayed result offers you the choice of whether your display indicates % moisture, % solids, % regain,
grams or custom unit during the measurement process.
TEST PARAMETERS
TEST ID:
PROFILE:
DRY TEMP:
SWITCH OFF:
RESULT:
CUSTOM:
-DEFAULT-
STANDARD
125C
AFREE
MOISTURE%
OFF
Using the arrow buttons, scroll to RESULT and press the Enter button.
% MOISTURE
EXIT WITHOUT SAVING
TEST PARAMETERS
Using the arrow buttons, scroll to either % MOISTURE, % SOLIDS, %
REGAIN, GRAMS or CUSTOM UNITS and press the Enter button.
NOTES: 1. If you change the displayed result during a test, the final
result and the printed results will be in displayed units.
2.
When CUSTOM is selected, the parameters for custom units
are changed in paragraph 6.7.