For menu items with numeric settings such as Capacity, the current

setting is displayed with all digits flashing.

Press the No button to begin editing.

The first digit is displayed flashing.

Press the No button to increment the digit or press the Yes button to

accept the digit and move to the next digit.


Repeat this process for all digits.

Press the Yes button when the last digit has been set.
The new setting is displayed with all digits flashing. Press the Yes

button to accept the setting or press the No button to resume editing.
This method also applies to setting Checkweigh under and over

targets.
For End menu items, pressing the Yes button advances to the next

menu, while pressing the No button returns to the top of the current

menu.

4.2 Calibration Menu

Enter this menu to perform calibrations.

Initial Calibration

When the scale is operated for the first time, a span calibration is recommended to ensure

accurate weighing results. Before performing the calibration, be sure to have the appropriate

calibration weights as listed in table 4-1. Ensure that the LFT switch/calibration lock is set to

unlocked position. See figure 5-1.
Or adjust the GEO setting according to your location (see table 4-2).
Procedure:
Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released, the display

will show [C.A.L]. Press Yes to accept. [SpaN] will then be shown. Press Yes to begin the span

calibration. [ 0

kg

] will be displayed. Press Yes to accept. [--C--] will be displayed while zero

reading is stored. Next, the display shows the calibration weight value. Place the specified

calibration mass on the pan. Press Yes to accept the weight or No to select an alternate weight.

[--C--] will be displayed while the reading is stored. The display will show [done] if the calibration

was successful. The scale returns to the previous application mode and is ready for use.

TABLE 4-1

Required Span Calibration Mass (sold separately)

Max

Mass

*

Max

Mass

*

1500g

1.5kg / 3lb

15000g

15kg / 30lb

3000g

3kg / 6lb

30000g

30kg / 60lb

6000g

6kg / 15lb

Note:

When active unit is g or kg, the calibrating unit will be in kg.

When active unit is lb, oz or lb:oz, the calibrating unit will be in lb.

For linearity calibration, the calibration Mass is fixed.

Mid-point is always half of full capacity.