EN-34

Ranger™ 3000

Series

7.5 Output Format

Weight string print format:
Field

Weight space Unit

space Stability space G/N

Space Term. Char(s)

Length 9

1

5

1

1

1

NET

1


Each field is followed by a single delimiting space (ASCII: 32)
Definitions:
Weight

- Up to 9 characters, right justified, - at immediate left of most significant character

(if negative).

Unit

- Up to 5 characters, left justified. If the Unit in the Print Content menu was set to

OFF, the unit will be removed in the weight string and replaced by spaces.

Stability

- “?” character is printed if not stable. If weight is stable neither “?” or following

space is printed.

G/N

- “NET” printed if weight is net weight, ‘G’ or nothing printed if weight is a gross

weight.

Terminating Character(s) - terminating character(s) printed depending on FEED menu setting.

7.6 Compliance

Compliance to the following standards is indicated by the corresponding mark on the product.

Mark

Standard

This product conforms to the EMC Directive 2004/108/EC, the Low

Voltage Directive 2006/95/EC and the Non-Automatic Weighing

Instrument Directive 2009/23/EC.

The Declaration of Conformity is available online at

europe.ohaus.com/europe/en/home/support/compliance.aspx.

AS/NZS CISPR 11

CAN/CSA-C22.2 No. 61010-1-12, UL Std. No. 61010-1 (3

rd

edition)


Important notice for verified weighing instruments

Weighing Instruments verified at the place of manufacture bear one of the

preceding marks on the packing label and the green ‘M’ (metrology) sticker on the

descriptive data plate. They may be put into service immediately.

Weighing Instruments to be verified in two stages have no green ‘M’ (metrology) on

the descriptive data plate and bear one of the preceding identification marks on the

packing label. The second stage of the initial verification must be carried out by an

authorized and certified service organization established within the European

Community or by the National Notified Body.

The first stage of the initial verification has been carried out at the manufacturers work. It

comprises all tests according to the adopted European standard EN 45501:1992, paragraph

8.2.2.

If national regulations limit the validity period of the verification, the user of the weighing

instrument must strictly observe the re-verification period and inform the respective weights

and measures authorities.