Specifications – Velleman DVM851 User Manual

Page 6

Advertising
background image

DVM851

01.08.2012

6

condensation is to be expected.(home and office environments fall under this
category)

Pollution degree 3 Conductive pollution occurs, or dry nonconductive pollution occurs that becomes

conductive due to condensation that is to be expected.
(industrial environments and environments exposed to outside air
contact with precipitation)

Pollution degree 4 The pollution generates persistent conductivity caused by conductive dust or by rain

or snow. (exposed outdoor environments and environments where high humidity
levels or high concentrations of fine particles occur)

Warning:
This device was designed in accordance with EN 61010-1 pollution degree 2. This implies that certain
restrictions in use apply that are related to pollution which can occur within the environment of use.
Refer to the table above.

This device is only suitable for measurements in Pollution degree class 2 environments.

9.

Specifications

This device is not calibrated when purchased!
Regulations concerning environment of use:

Use this meter only for measurements in CAT I and CAT II environments (see §7)

Use this meter only in a pollution degree 2 environment (see §8)

Ideal working conditions include:

Temperature: 18 to 28°C (64°F to 82°F)

relative humidity: max. 80%

altitude: max. 2000m (6560ft)

voltage

max. 600V

Fuse protection

F 500mA/250V, F10A/250V

Power

9V alkaline battery 6LR61

Display

LCD, 1999 counts

sampling rate

2-3/sec.

Measuring method

Dual-slope integration A/D converter

Overrange indication

Only figure "1" on the display

Polarity indication

"-" displayed for negative polarity

Operating temperature

0 to 40°C

Storage temperature

-10°C to 50°C

Low battery indication

appears on the display

Dimensions

138 x 69 x 31mm

Weight

Approx. 142g

9.1 DC VOLTAGE

Do not measure circuits that may contain voltages > 600V

Range

Resolution

Accuracy

200mV

100µV

±

0.5% of rdg

2V

1mV

±

0.8% of rdg

20V

10mV

200V

100mV

600V

1V

±

1.0% of rdg

9.2 DC CURRENT

Do not measure current in circuits with voltages > 250V

Range

Resolution

Accuracy

200µA

0.1µA

±

1% of rdg

2mA

1µA

20mA

10µA

200mA

100µA

±

1.5% of rdg

10A

10mA

±

3% of rdg

Overload protection by fuse: 200mA range F 500mA/250V, 10A range F10A/250V.

©Velleman nv

condensation is to be expected.(home and office environments fall under this

tive pollution occurs, or dry nonconductive pollution occurs that becomes

(industrial environments and environments exposed to outside air - but not in

he pollution generates persistent conductivity caused by conductive dust or by rain

or snow. (exposed outdoor environments and environments where high humidity

. This implies that certain

restrictions in use apply that are related to pollution which can occur within the environment of use.

lution degree class 2 environments.

Use this meter only for measurements in CAT I and CAT II environments (see §7)

10A/250V

6LR61

slope integration A/D converter

Only figure "1" on the display

" displayed for negative polarity

appears on the display

Accuracy

0.5% of rdg

±

2 digits

0.8% of rdg

±

2 digits

1.0% of rdg

±

2 digits

Accuracy

1% of rdg

±

2 digits

1.5% of rdg

±

2 digits

3% of rdg

±

2 digits

250V.