10 connecting the voltage supply, 1 diq/s 182 xt-4-pr (line power version), 10 connecting the voltage supply -35 – YSI DIQ/S 182 XT-4-PR User Manual

Page 49: 1 diq/s 182 xt-4-pr (line power version) -35

Advertising
background image

System 182 XT-4

Installation

3 - 35

ba76027e01

02/2012

3.10

Connecting the voltage supply

Note
The following two paragraphs describe how to connect both models of
the DIQ/S 182 XT-4-PR to the voltage supply. How to connect
additional power supply modules is described in the operating manual
of the respective power supply module.

3.10.1 DIQ/S 182 XT-4-PR (line power version)

Warning

If the power supply is incorrectly connected, it may represent a
danger to life from electric shock. Pay attention to the following
points during installation:
The DIQ/S 182 XT-4-PR must only be connected by a qualified

electrician.

The DIQ/S 182 XT-4-PR must only be connected in a voltage-

free condition.

The power supply must fulfill the specifications given on the

nameplate and in chapter 10 T

ECHNICAL

DATA

.

A switch or circuit breaker must be provided in the building

installation as an interrupt facility for the DIQ/S 182 XT-4-PR.
The interrupt facility must be

– installed in the vicinity of the DIQ/S 182 XT-4-PR, easily

accessible by the user and

– marked as an interrupt facility for the DIQ/S 182 XT-4-PR.

After it has been installed, the DIQ/S 182 XT-4-PR may only be

opened if the mains voltage has been switched off beforehand.

Materials required

 Wire end sleeves, suitable for the power line, with suitable crimping

tool

 1 x cable gland, matched to the cable diameter (see section 3.9 on

page 3-33).

Tools

 Cable stripping knife

 Wire stripper

 Phillips screwdriver

 Small screwdriver.

Preparing the power

cable

1

Cut off the cable to the required length.

2

Strip the cable insulation for approx. 45 mm.

Advertising