System 182 XT-4

Installation

3 - 19

ba76027e01

02/2012

Materials required

 1 x SNCIQ or SNCIQ/UG connection cable (see

chapter 12 A

CCESSORIES

AND

OPTIONS

)

 Wire end sleeves for 0.75 mm2 wire cross-section with matching

crimping tool

 1 x cable gland, matched to the cable diameter (see section 3.9 on

page 3-33).

Tools

 Cable stripping knife

 Wire stripper

 Phillips screwdriver

 Small screwdriver.

Preparing the cable

ends

1

Cut off the cable to the required length.

2

Remove approx. 45 mm of cable insulation (in the case of the
SNCIQ/UG earth cable, remove both the inner and outer
insulation).

3

Only for the SNCIQ/UG earth cable:
strip the outer insulation for a further 35 mm.

4

Shorten the exposed shielding braid up to the cable sheath.

5

Shorten the two fillers (plastic inlays) up to the cable sheath.

6

Bare the red and green wires and fit them with wire end
sleeves.

7

Fit the filler stranded wire with a wire end sleeve.