YSI DIQ/S 182 XT-4-PR User Manual
Page 135
System 182 XT-4
What to do if ...
9 - 5
ba76027e01
02/2012
Operating sequence in
case 2
1
Connect the IQ sensor.
2
Change to the measured value display with
m
. The
component database is updated. The following display
appears (example):
3
Select the required option with
w
and confirm with
g
.
– If Add as new sensor was selected, the system changes
directly to the measured value display. As soon as the
IQ sensor is ready for operation, it delivers a measured
value.
– If Substitute sensor was selected, a list with the assignable
sensors or inactive datasets appears.
4
Select the required sensor with
w
and confirm with
g
. The
system changes to the measured value display. The IQ sensor
takes over all settings of the inactive dataset. As soon as the
IQ sensor is ready for operation, it delivers a measured value.