4.2.5

Reactivating previous calibration data

The FDO

®

70x IQ (SW) enables you to reactivate the last valid user

calibration or the factory calibration. Thus you can immediately go on
measuring if a calibration procedure failed or you suspect that the cali-
bration conditions were not optimally met.

Note
Reactivating old calibration data is only a temporary solution. Please
take into account that the sensor might thus possibly provide incorrect
measured values. Please make sure the sensor works correctly by car-
rying out a function check and/or recalibration.

Reactivating calibration

data

1

Open the setting table (see section 3.4).

2

In the Calibration menu item, select the User calibration or Fac-
tory calibration
setting and then quit the setting table with Save
and quit
.