5 technical data of the sc-fdo® 700, Technical data of the sc-fdo – Xylem IQ SensorNet FDO 70x IQ (SW) User Manual

Page 38

Advertising
background image

Technical data

FDO

®

70x IQ (SW)

7 - 4

ba76014e01

02/2012

7.5

Technical data of the SC-FDO

®

700

Measuring ranges and

resolutions

D. O. partial pressure 0 ... 400 hPa

Measurement accuracy

Repeatability

± 0.05 mg/L (ppm) according to EN ISO 15839

Response time

t

90

(90 % of the final display value after) < 150 s according to

EN ISO 15839
t

95

(95 % of the final display value after) < 200 s according to

EN ISO 15839

Minimum approach flow

No incident flow required

Interferences

Non-sensitive to against hydrogen sulfide, chlorine, and ionogenic
substances

Allowed temperature

range

Allowed pH range of the

test sample

4 ... 12

Calibration procedure

Factory calibration by means of IQMC (Intelligent Membrane Calibra-
tion) procedure. User calibration possible in water vapor-saturated air
or air-saturated water.

Material

PMMA, PVC and silicone

Working life

2 years guaranteed with authorized use

Measuring
mode

Adjustable measuring range

Resolution

D. O. concentra-
tion

0 ... 20.00 mg/L
0 ... 20.00 ppm

0.01 mg/L
0.01 ppm

D. O. saturation

0 ... 200.0 %

0.1 %

In the range

Accuracy

< 1 mg/L (ppm)

± 0.05 mg/L (ppm)

> 1 mg/L (ppm)

± 0.1 mg/L (ppm)

Measuring medium

-5 °C ... + 50 °C (23 ... 122 °F)

Storage/transport

- 25 °C ... + 50 °C (-33 ... 122 °F)

Advertising