70x IQ (SW)

What to do if...

6 - 1

ba76014e01

02/2012

6

What to do if...

Sensor does not appear

in the measured value

display and list of

sensors

Measured value

implausible

Measured value

fluctuating heavily

Cause

Remedy

– Sensor cap not mounted or

defective

– Sensor cap (see section 5.3)

– Replace a defective sensor

cap as necessary

Cause

Remedy

– Coating on sensor cap

– Clean the outside of the sen-

sor (see section 5.4.1)

– Service life of the sensor cap

over

– Check the sensor (see sec-

tion 4.2.2)

– Replace the sensor cap as

necessary (see section 5.3)

– Dirt inside the sensor cap and

sensor head.

– Fixing ring not properly tight-

ened.

– Sensor cap untight or defec-

tive.

– Dismantle sensor cap

– Clean the insides of the sen-

sor cap and sensor head (see
section 5.4.2)

– Mount the sensor cap cor-

rectly and tighten the fixing
ring as far as it will go (see
section 5.3)

– Replace a defective sensor

cap as necessary

Cause

Remedy

– Coating on sensor cap

– Clean the outside of the sen-

sor (see section 5.4.1)

– Service life of the sensor cap

over

– See point "Measured value

implausible"

– Dirt inside the sensor cap and

sensor head.

– Fixing ring not properly tight-

ened.

– Sensor cap untight or defec-

tive.

– See point "Measured value

implausible"