Info messages

Message code

Message text

IC133x

Sensor has been successfully calibrated
* For calibration data, see calibration history

IC333x

Factory calibration has been activated. Make sure the sensor operates
correctly.

IC433x

Last valid user calibration has been activated. Make sure the sensor
operates correctly.

IC533x

Invalid user calibration has been replaced by last valid user calibration.
Caution! Wrong measured values possible. Carry out a new successful
user calibration to make sure the sensor operates correctlyes correctly.

IC633x

The invalid user calibration was replaced by the factory calibration.
Caution! Wrong measured values possible. Check whether the sensor
operates correctly or carry out a new successful user calibration user
calibration.

IC733x

This message displays the result of the last check (see section 4.2.2)

IS133x

This message displays the information stored in the sensor cap
(see section 1.2)