MIQ/IC2

Installation

3 - 5

ba76034e01

01/2012

Connecting external

meters via a power sup-

ply/isolator

The current outputs of external meters can be directly connected to the
current inputs of the MIQ/IC2 module via a power supply/isolator. Con-
necting via a power supply/isolator is necessary for measurements in
potentially explosive areas, for example.

If the connection data of the power supply/isolators is suitable, it is pos-
sible to supply up to two power supply/isolators with power via the feed-
in connector on the MIQ/IC2 module.

Fig. 3-3

Connection diagram for the connection of external meters situated in

potentially explosive areas to the MIQ/IC2 module

Warning

In potentially explosive areas, instruments may be installed, com-
missioned and operated by specialist electricians only, and ac-
cording to the relevant regulations and the instructions in the
relevant operating manuals.

IQ Sensor Net

2 x Power supply/isolator

Potentially
explosive area,
zone 1 / zone 2

4 … 20 mA

4 … 20 mA

4 … 20 mA

4 … 20 mA

Feed-in

1
5

7
3

4 8

11
12

11

12

-

+

-

+

-

+

-

+

Meter 1

Meter 2

-

+

5

-

+

1

7
3

4 8

MIQ/IC2

- +

X7 X8

X9

X10

X11
X12

-

+

-

+