MIQ/IC2

What to do if ...

5 - 1

ba76034e01

01/2012

5

What to do if ...

No measured value

Measurement provides

implausible measured

values

Display of OFL

Cause

Remedy

– MIQ/IC2 not connected

– Connect the MIQ/IC2

– Unknown

– Look in the log book

Cause

Remedy

– Unsuitable settings of:

Measuring range,

Display value (0/4 mA),

Display value (20 mA)

– Adjust the settings in the setting

menu of the MIQ/IC2

– Current value at the current

input or output of an instru-
ment (e.g. measuring trans-
mitter, power supply/isolator,
MIQ/IC2 etc.) does not agree
with the nominal value

– Change the settings of the

external meters as necessary

– Adjust the settings in the setting

menu of the MIQ/IC2

– Calibration is not up-to-date

– Calibrate the external meter

Cause

Remedy

– Allowed signal range excee-

ded or undercut

– Operate the 0/4 ... 20 mA cur-

rent inputs of the MIQ/IC2 in the
allowed current range only (see
chapter 7 T

ECHNICAL

DATA

)