7 technical data, Technical data -1, Chapter 7 t – YSI IQ SensorNet MIQ/IC2 Module User Manual

Page 23: Echnical, Data, Ee chapter 7 t, 7technical data

Advertising
background image

MIQ/IC2

Technical data

7 - 1

ba76034e01

01/2012

7

Technical data

Note
General technical data on MIQ modules are given in the T

ECHNICAL

DATA

chapter of the IQ S

ENSOR

N

ET

system operating manual.

Electrical data

Instrument safety

Terminal connections

Nominal voltage

Max. 24 VDC via the IQ S

ENSOR

N

ET

(for

details, see the T

ECHNICAL

DATA

chapter of

the IQ S

ENSOR

N

ET

system operating

manual)

Power consumption

0.2 ... 4.6 W

0.2 W: without supplying any power supply/
isolator

2.4 W: with one power supply/isolator

4.6 W: with two power supply/isolators

Protective class

III

Applicable norms

– EN 61010-1
– UL 3111-1
– CAN/CSA C22.2 No. 1010.1

IQ S

ENSOR

N

ET

connec-

tions

2
Additional connectable SENSORNET termi-
nator (terminating resistor)

Current inputs
(0/4 ... 20 mA)

2

Connector for power
supply/isolator

1

Terminal type

Screw-type terminal strip, accessible by
opening the lid

Terminal ranges

Solid wires:

0.2 ... 4.0 mm

2

AWG 24 ... 12

Flexible wires:

0.2 ... 2.5 mm

2

Line cross-section of ca-
bles carrying mains volt-
age

Europe:

1.5 ... 4.0 mm

2

USA:

AWG 14 ... 12

Cable feeds

4 cable glands M16 x 1.5 on the underside
of the module