8 programs, Programs, Section 4.5.8) – YSI pHotoFlex STD Colorimeter User Manual
Page 43
pHotoFlex
®
STD
Operation
43
ba76122e01
02/2013
If
iCalibration error! was displayed as the calibration result, a note au-
tomatically reminds you of another zero adjustment before the next
measurement.
Measuring is not possible without a valid zero adjustment.
4.5.8
Programs
Displaying program
data
You can view the most important data of all methods.
The method data is ordered according to the program number.
4
Insert the cell (see section 4.2).
5
Set another cell with <> <> and <START/ENTER> as nec-
essary.
6
Start the measurement of the zero adjustment with <START/
ENTER>.
After measuring, the result of the zero adjustment is displayed
and stored.
iZero adjust. successful! (successful zero adjustment) or
iCalibration error! (zero adjustment not successful) is dis-
played as the result.
The zero adjustment is completed.
7
Confirm the result with <START/ENTER>.
The zero adjustment is completed.
The meter is ready to measure.
iInsert zero
cell (dist. water)
Cell
16 mm
Start measurement
Zero adjust.
1
Open the Configuration / Photometry / Programs menu.
The display shows the most important data of the selected pro-
gram.