4.5.6
Absorbance / % Transmission, measuring
The transmission measurement is not described separately in the fol-
lowing example as it operates in exactly the same way as the absor-
bance measurement. The result of the measurement is displayed in %
Transmission.
1
Press the <M> several times as necessary until the measured
parameter Absorbance (or % Transmission) is selected.
2
Select the cell diameter with <> <> and confirm with
<START/ENTER>.
3
Select the wavelength with <> <> and confirm with
<START/ENTER>.
4
Clean the cell (see section 5.2.2).
5
Insert the cell (see section 4.2).
6
Start the measurement with <START/ENTER>.
The measurement result is displayed when the measurement
is finished.
Photometry \ Absorbance
iSelect cell
with
▲
▼
16 mm
610 nm
01.12.11 15:12
Photometry \ Absorbance
iInsert sample
iStart measurement
with <START>
16 mm
610 nm
01.12.11 15:12