YSI pHotoFlex STD Colorimeter User Manual

Page 59

Advertising
background image

pHotoFlex

®

STD

Operation

59

ba76122e01

02/2013

The individual data sections of the command line are separated by
commas. The dot "." has to be used as a decimal separator within a
data section.

If there is already a program stored under the selected number, the pro-
gramming is canceled. If you wish to store the program under the se-
lected number, you have to erase the program stored under the
number first.

After a successful transmission, the terminal program writes "!>".
After a failed transmission, the terminal program writes "!>".

Erasing user-defined

programs

To erase user-defined programs, enter a command line according to
the following scheme in the terminal program:

After a successful transmission, the requested data is displayed on the
terminal.
After a failed transmission, the terminal program writes "!>".

Reading out user-

defined programs

To read out user-defined programs, enter a command line according to
the following scheme in the terminal program:

5

Enter the command line.

6

Finish the command line with Enter.

The data is transmitted to the pHotoFlex

®

STD.

Erase all user-defined
programs

Erase one user-
defined program

Command line

U.520

U.521#1,Program No

Example

U.520

U.521#1,900

7

Enter the command line.

8

Finish the command line with Enter.

The data is transmitted to the pHotoFlex

®

STD.

The requested data is displayed on the terminal as the result.

Read out all user-
defined programs

Read out one user-
defined program

Command line

U.510

U.511#1,Program No

Example

U.510

U.511#1,900

9

Enter the command line.

Advertising