YSI pHotoFlex STD Colorimeter User Manual
Page 40
Operation
pHotoFlex
®
STD
40
ba76122e01
02/2013
If a standard adjustment is stored for a program, the standard adjust-
ment is automatically used for measurement. The usage of the stan-
dard adjustment is documented together with the measured value and
indicated in the measured value display with [Cal].
6
Enter the nominal value of the standard.
Note:
Enter the decimal separator with <START/ENTER>.
7
Confirm the entered nominal value with <START/ENTER>.
8
Insert a cell with standard (see section 4.2).
9
Start the measurement of the standard with <START/ENTER>.
After measuring, the result of the standard adjustment is dis-
played and stored.
As the result, the measured value and the adjustment (in %) or
Error is displayed.
10
Confirm the result with <START/ENTER>.
The standard adjustment is completed.
The meter is ready to measure.
or:
Discard the result with <ESC>.
Subsequently, carry out a new standard adjustment.
mg/l Cu
_
Enter nominal value
i304: Cu-1 TP
Cell
i = 10 mm
i 0.600 mg/l Cu:
2.000 (93.2%)
Accept
Measure standard