4.11

Administrating user-defined methods

4.11.1 Options for method administration

User-defined programs can be

 entered
 read out
 erased.

To store user-defined programs (methods) in the meter, determine a
calibration line for your method yourself and transmit the data of this
calibration line to the pHotoFlex

®

STD.

Administration of the method data can be carried out in the LSdata soft-
ware (see LSdata operating manual) or with the aid of a terminal pro-
gram ( see section 4.11.2).

You need the following data in any case:

Data

Possible entries/examples

Program No:

900 ... 949

Model:

Any name (max. 5 characters)

Cell [mm]:

16 or 28

Wave length [nm]:

436, 517, 557, 594, 610 or 690

MRB (measuring range start):

e.g. 0.05

MRE (measuring range end):

e.g. 8.00

E0 (reagent blank value):

e.g. 0.0

Slope: e.g.

1.0

Resolution:

0.0001, 0.001, 0.01, 0.1, 1, 2, 5,
10 or 100

Citation form:

e.g. PO4-P (max. 9 characters)

Unit:

e.g. mg/l (max. 7 characters)

Blank required:
(No/Yes)

0 or 1

Standard adjustment possible:
(No/Yes)

0 or 1

Standard adjustment required:
(No/Yes)

0 or 1