6 calibration data, Calibration data – YSI TruLab 1310P User Manual

Page 32

Advertising
background image

pH value

TruLab pH 1310

32

ba76156e01

03/2014

5.2.6 Calibration data

Displaying the cali-

bration data

The calibration record of the last calibration is to be found under the menu item,
<F1> / [Menu]Calibration / Calibration record. To open it, press the <CAL__>
key in the measured value display.

Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with <F2>[USB output].

Displaying the cali-

bration data storage

The calibration records of the last calibrations (up to 10) are available in the
menu, <F1>/[Menu] / Calibration / Calibration data storage and in the menu,
<F1__>/[Menu] / Storage & config/Data storage / Calibration data storage.

Calibration evalua-

tion

After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.

The calibration record is automatically transmitted to the interface
after calibrating.

Menu item

Setting/func-
tion

Description

Calibration /
Calibration data stor-
age
/Display

or

Data storage /
Calibration data stor-
age
/Display

-

Displays the calibration record.

Further options:

Scroll through the calibration
records with <><>.

Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].

Quit the display with <F1>/
[Back] or <MENU/ENTER>.

Switch directly to the mea-
sured value display with <M>.

Calibration /

Calibration data stor-
age
/

Output to USB
or

Data storage /
Calibration data stor-
age
/Output to USB

-

Outputs the calibration records to
the interface.

Advertising
This manual is related to the following products: