3 calibration interval, 2 settings for orp measurements, Calibration interval – YSI TruLab 1310P User Manual

Page 42: Settings for orp measurements

Advertising
background image

Settings

TruLab pH 1310

42

ba76156e01

03/2014

*

Brand names or trade names are trademarks of their respective owners pro-
tected by law.

7.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol.

The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.

Setting the calibra-

tion interval

The calibration interval is set to 7 days in the factory.

You can change the interval (1 ... 999 days):

7.2

Settings for ORP measurements

The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required parame-

23

Reagecon 25 *

2.000
4.000
7.000
10.000
13.000

25 °C

24

Chemsolute *

2,000
4,000
7,000
10,000

20 °C

25

USABlueBook

*

4,000
7,000
10,000

25 °C

No.

Buffer set *

pH values

at

The buffers are selected in the menu, pH / <F1>/[Menu] / Calibra-
tion
/ Buffer (see 7.1.1 C

HANGING

THE

SETTINGS

FOR

P

H

MEASURE

-

MENTS

,

PAGE

38).

To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.

1.

Open the menu for measurement settings with [Menu]<F1>.

2.

In the Calibration / Calibration interval menu, set the calibration interval
with <><>.

3.

Confirm the setting with <ENTER>.

4.

Quit the menu with <M>.