Ph 1 – YSI TruLab 1310P User Manual

Page 26

Advertising
background image

pH value

TruLab pH 1310

26

ba76156e01

03/2014

Continuing with two-

point calibration

5.

Thoroughly rinse the electrode with deionized water.

6.

Immerse the electrode in the first buffer solution.

7.

When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.

8.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

9.

Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.

The calibration display for the next buffer appears (voltage display).

10. If necessary, finish the calibration procedure as a single-point calibration

with <M>.
The calibration record is displayed.

For single-point calibration, the instrument uses the Nernst slope

(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.

pH

1

-180.0

24.8 °C

AutoCal YSI

01.02.2014

08:00

Buffer

mV

pH

-180.0

24.8 °C

AutoCal YSI

10.00

01.02.2014

08:00

AR

Buffer

mV

11. Thoroughly rinse the electrode with deionized water.

12. Immerse the electrode in the second buffer solution.

13. When measuring without temperature sensor:

Measure the temperature of the buffer and enter it with <><>.