YSI TruLab 1310P User Manual

Page 27

Advertising
background image

TruLab pH 1310

pH value

ba76156e01

03/2014

27

Continuing with

three- to five-point

calibration

14. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

15. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with
<ENTER>.

The calibration display for the next buffer appears (voltage display).

16. If necessary, finish the calibration procedure as a two-point calibration

with <M>.
The calibration record is displayed.

pH

0.0

24.8 °C

AutoCal YSI

7.000

01.02.2014

08:00

AR

Buffer

mV

17. Thoroughly rinse the electrode with deionized water.

18. Immerse the electrode in the next buffer solution.

19. When measuring without temperature sensor:

Measure the temperature of the buffer and enter it with <><>.

20. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

pH

180.0

24.8 °C

AutoCal YSI

4.00

01.02.2014

08:00

AR

Buffer

mV