1 resetting the measurement settings, 2 resetting the system settings, Resetting the measurement settings – YSI TruLab 1310P User Manual

Page 45: Resetting the system settings, Section 7.4.1 r, Esetting, Measure, Ment, Settings, Measurement

Advertising
background image

TruLab pH 1310

Settings

ba76156e01

03/2014

45

7.4.1 Resetting the measurement settings

The following settings for pH measurements are reset to the default settings
with the Reset function:

The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it, press the <F1>/[Menu] or
<MENU/ENTER> key in the measured value display.

7.4.2 Resetting the system settings

The following system settings can be reset to the default condition:

The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.

Setting

Default settings

Buffer

YSI

Cal. interval

7 d

Unit for slope

mV/pH

Measured parameter

pH

Resolution pH

0.001

Resolution mV

0.1

Asymmetry

0 mV

Slope

-59.2 mV

Man. temperature

25 °C

One point calibration

No

Stability control

On

Temperature unit

°C

Setting

Default settings

Language

English

Beep

On

Baud rate

4800 Baud

Output format

ASCII

Contrast

50 %

Illumination

Auto

Switchoff time

1 h