5 what to do if... / printer – YSI TruLab 1310P User Manual

Page 57

Advertising
background image

TruLab pH 1310

Printer (only TruLab pH 1310P)

ba76156e01

03/2014

57

10.5 What to do if... / printer

Integrated printer

does not print

Printer operating -

paper not being

printed

Integrated printer

prints automatically

Cause

Remedy

– Printer switched off

(LED flashing)

– Switch on the printer

(LED illuminated)

– No power pack connected

– Connect the power pack

– USB cable connected

– Disconnect the USB cable from the

meter

– Function "automatic storing at

intervals" with long intervals is
switched on

– Switch off the function (see section

8.2 A

UTOMATIC

STORAGE

AT

INTER

-

VALS

, page 48)

– No paper available

– Insert a roll of paper

Cause

Remedy

– Paper inserted the wrong way up – Turn the roll of paper around and

insert it with the other side upwards

Cause

Remedy

– The function "Automatic storage

at intervals" or "Transmit data
automatic at intervals" is
switched on

– Switch off the function (see section

8.2 A

UTOMATIC

STORAGE

AT

INTER

-

VALS

, page 48 or

section 9.1 O

PTIONS

FOR

DATA

TRANSMISSION

, page 53)

Advertising