2 general information – YSI TruLab 1310P User Manual

Page 62

Advertising
background image

What to do if...

TruLab pH 1310

62

ba76156e01

03/2014

12.2 General information

Sensor symbol

flashes

Display

Instrument does not

react to keystroke

You want to know

which software

version is in the

meter

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Batteries almost empty

– Replace the batteries (see

section 11.1 M

AINTENANCE

,

page 58)

Cause

Remedy

– Operating condition undefined or

EMC load unallowed

– Processor reset:

Press the <ENTER> and <On/
Off>
key simultaneously

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

Open the menu, <F1_>/
[Menu] / Storage & config /
SystemService information.
The instrument data are dis-
played.

Advertising
This manual is related to the following products: