Page 62
Advertising
What to do if...
TruLab pH 1310
62
ba76156e01
03/2014
12.2 General information
Sensor symbol
flashes
Display
Instrument does not
react to keystroke
You want to know
which software
version is in the
meter
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see
Cause
Remedy
– Operating condition undefined or
EMC load unallowed
– Processor reset:
Press the <ENTER> and <On/
Off> key simultaneously
Cause
Remedy
– E. g., a question by the service
department
– Switch on the meter.
Open the menu, <F1_>/
[Menu] / Storage & config /
SystemService information.
The instrument data are dis-
played.