13 Technical data

13.1 Measuring ranges, resolution, accuracy

Measuring ranges,

resolution

Parameter

Measuring range

Resolution

pH

-2.0 ... +20.0

0.1

-2.00 ... +20.00

0.01

-2.000 ... +19.999

0.001

U [mV]

-1200.0 ... +1200.0

0.1

-2500 ... +2500

1

T [°C]

-5.0 ... +105.0

0.1

T [°F]

23.0 ... +221.0

0.1

Manual

temperature input

Parameter

Range

Increment

T

manual

[°C]

-25 ... +130

1

T

manual

[°F]

-13 ... +266

1

Accuracy (± 1 digit)

Parameter

Accuracy

Temperature of the test
sample

pH / range *

-2.0 ... +20.0

± 0.1

+15 °C ... +35 °C

-2.00 ... +20.00

± 0.01

+15 °C ... +35 °C

-2.000 ... +19.999

± 0.005

+15 °C ... +35 °C

U [mV] / range

-2500 ... +2500

± 1

+15 °C ... +35 °C

-1200.0 ... +1200.0

± 0.3

+15 °C ... +35 °C

T [°C] / temperature sensor

NTC 30

± 0.1

PT 1000

± 0.1

* when measuring in a range of ± 2 pH around a calibration point

The accuracy values specified here apply exclusively to the meter.
The accuracy of the electrodes and buffer solutions has to be taken
into account additionally.