Xylem ProLab 3000 User Manual
Page 100

Ion concentration
ProLab 3000
100
ba75645e03
12/2012
– Address of TISAB/ISA :
(1...3)
– Address of standard/sample :
(1...3)
– (Address of blank value:)
(1...3)
Each burrete must have a distinct
address (1...3) so it can be indi-
vidually addressed.
INFO: The address of the burette
must also be set on the burette
(see operating manual of the
burette).
– Stability control
– High ()
– Medium
– Low
Selection of criteria for the stablity
control (see page 98).
– ISE methods
– Standard addition...
– Double standard addition
– Standard subtraction...
– Sample addition...
– Sample subtraction...
– Blank value addition...
Selection of the available measur-
ing methods.
– Alternative TP
()
() : Alternative temperature sen-
sor is used
() : no alternative temperature
sensor is used
– Setting of man. temp...
-35 ... +150 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Menu item
Setting
Description
Measuring