Ion concentration

95

ba75645e03

12/2012

9

Ion concentration

9.1

General information

Temperature

measurement with ion

selective measurements

The temperature value is documented with all measurements.

You have the following options to measure the temperature:

 Automatic temperature measurement with

– the temperature sensor integrated in the electrode (NTC 30 or

Pt 1000).

– the integrated temperature sensor of another electrode that is

connected and immersed in the sample at the same time.

– an external temperature sensor (NTC 30 or Pt 1000).

 Manual determination and input of the temperature.

The display of the temperature indicates the active temperature mea-
suring mode:

Operation with burettes

Measurements with ISE methods can be automated by controlling
burettes with the meter:

 Connect the burette to the RS232 interface of the ProLab 3000

 If necessary, interlink the burettes

(see also operating manual of the burette).

 Activate the burette control on the ProLab 3000:

Measuring / Burettes settings... / Activate burettes ()

(see section 9.2.3).

 Set the address of the burette on the meter and on the burette

(see page 100 and operating manual of the burette).

Note
Incorrect calibration of ion sensitive electrodes will result in incorrect
measured values. Calibrate regularly before measuring.

Temperature

sensor

Resolution of the
temp. display

Mode

yes

0.1 °C

Automatic with tempera-
ture sensor

-

1 °C

Manual

Note
More information on automation with burette control is available on the
Internet under www.si-analytics.com.