8 automatic stability control, Automatic stability control – Xylem ProLab 3000 User Manual
Page 63

ProLab 3000
System functions
63
ba75645e03
12/2012
6.8
Automatic stability control
The Autom. stability control function continuously checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values.
You can activate or switch off the Autom. stability control function
(menu, Autom. stability controlSystem).
The Autom. stability control function is carried out:
as soon as the measured value is outside the allowed stability range
when you switch over between the measured parameters with
<MODE>.
Whether or not a measured value is stable can be taken from its color
labeling.
In the delivery condition, the SC stable / SC not stable conditions are
assigned to the following colors:
Color
Meaning
red
SC not stable
Stability criteria not met
green
SC stable
Stability criteria met
SC stable
SC not stable