Ion concentration
ProLab 3000
104
ba75645e03
12/2012
9.3.1
Carrying out calibration
Use two to nine standard solutions in ascending or descending order.
Preparatory activities
Perform the following preparatory activities when you want to calibrate:
Calibration
Note
When calibrating within the framework of a method measurement, it is
always a two-point calibration.
1
Connect an ion sensitive electrode to the meter.
2
Highlight the channel of the electrode in the measured value
display.
3
If necessary, select the ISE display with <MODE>.
4
If necessary, open the Measuring / ISE settings / Ion type menu
and select the ion type to be measured.
5
Keep the standard solutions ready.
6
If necessary, measure the temperature of the standard solu-
tions with a thermometer.
7
If necessary, change the unit of the measurement result and
calibration standards in the menu, Measuring/Units.
1
Perform the preparatory activities (see page 104).
2
Start the calibration with <CAL>.
The ISEX calibration window opens.