3 general errors – Xylem ProLab 3000 User Manual
Page 176

What to do if...
ProLab 3000
176
ba75645e03
12/2012
Warning [TempErr]
Warning [ISEErr]
Burette does not dose
15.3
General errors
Meter does not react to
keystroke
Time is lost
Cause
Remedy
– Temperature difference
between measurement and cali-
bration greater than 2 K.
– Adjust the temperature of
the test sample
Cause
Remedy
– Electrode voltage outside cali-
brated range
– Recalibrate
Cause
Remedy
– Address of the burette incor-
rectly set
– Enter the address of the
burette on the meter
– Enter the address of the
burette on the burette
(see operating manual of the
burette)
– RS232 interface of the burette
incorrectly set
– Configure the RS232 inter-
face of the burette
(see page 29 and operating
manual of the burette)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <On/Off> and
<SC> key simultaneously.
Cause
Remedy
– Buffer batteries empty
– Exchange buffer batteries
(see page 171)