Switching on the measuring instrument

pH 1970i

20

ba75675e01

01/2007

Switching on the measuring instrument

Measuring

Overview of the measuring modes:

Special functions:

AutoRead

(drift control)

The AutoRead function checks the stability of the measurement signal
(except for the measurement of the ORP voltage). Activate AutoRead with
<AR>. Press <RUN/ENTER> to start the AutoRead measurement. During
the AutoRead measurement, AR flashes on the display until a stable
measured value is reached. This can be terminated at any time taking over
the current measured value with <RUN/ENTER>.

Temperature

measurement

during pH

measurements

You can use pH electrodes with and without temperature sensors. The
measuring instrument recognizes which temperature sensors are connected
and switches automatically to the correct mode for the temperature
measurement (display TP).
Manual temperature input: Set the temperature value with <

> <

>.

z

Press the <ON/OFF> key.
The Display test appears briefly on the display.
After this, the measuring instrument automatically switches to the
measuring mode. The display shows the relevant measured value.

pH value

Redox-
voltage

mV