Saving data

25

ba75675e01

01/2007

Saving data

The measuring instrument has an internal data memory. It can store up to
800 datasets.

Saving data

manually

Press the <STO> key in the measuring mode (display No. with the number of
the next free memory location). Then, press <RUN/ENTER> and enter the ID
number with <

> <

>. Finish the save with <RUN/ENTER>. The instrument

changes to the measuring mode.

Saving data

automatically Int 1

The save interval (Int 1) determines the chronological interval between
automatic save processes.
To set up the save interval, press <STO> while pressing the <RUN/ENTER>
key (display Int 1) and set the interval with <

> <

> key. Then press <RUN/

ENTER>. The number of the available free memory locations is displayed.
Press <RUN/ENTER> once more and enter the ID number with <

> <

>.

Switch to the last active measuring mode with <RUN/ENTER>. The
automatic save is switched on (display Auto Store).

Outputting the data memory

You can output the data memory with the <RCL> key. By repeatedly pressing
the <RCL> key, you reach the following functions:

Start the output with <RUN/ENTER>.

Erasing the data storage

Switch off the measuring instrument. Press the <STO> key and keep it
depressed. Press the <ON/OFF> key. Confirm the erasing procedure with
<RUN/ENTER> or cancel the erasing procedure by pressing any other key.

Transmitting data

Transmitting data

manually

Press the <RUN/ENTER> key in the measuring mode.
This manually triggers a data transmission of the current measured value to
the serial interface at any time - independently of the selected intervals.

StO dISP

Output stored data on the display

StO SEr

Output stored data on the serial interface

CAL dISP

Output calibration data on the display

CAL SEr

Output calibration data on the serial interface