Measuring ranges, resolutions, accuracies – Xylem PH1970i PORTABLE PH METER User Manual

Page 30

Advertising
background image

Technical data

pH 1970i

30

ba75675e01

01/2007

Measuring ranges, resolutions, accuracies

* when measuring in a range of ± 2 pH around a calibration point

Analog output

(AK 323/S cable)

Automatic switchover when the recorder is connected by the cable,
AK 323/S.

pH output signal

-200 ... +1999 mV for the range
- 2.00 ... + 19.99

mV output signal

-1999 ... +1999 mV for the range
-1999 ... +1999 mV for the range

Accuracy

± 0.5 % of display value

Internal resistance

< 5 Ohm (current limited to max. 0.2 mA
output current)

Guidelines

and norms used

EMC

EC guideline 89/336/EEC
EN 61326-1:1998
EN 61000-3-2 A14:2000
EN 61000-3-3:1995
FCC Class A

Instrument safety

EC guideline 73/23/EEC
EN 61010-1 A2:1995

Climatic class

VDI/VDE 3540

IP protection

EN 60529:1991

Variable

Measuring
range

Resolution

Accuracy (± 1 Digit)

pH

- 2.00 ... + 19.99

0.01

± 0.01 *

U [mV]

- 199.9 ... + 199.9
- 1999 ... + 1999

0.1
1

± 0.5
± 1

T [°C]

- 5.0 ... + 105.0

0.1

± 0.1

Manual

temperature

input

Variable

Range

Increment

T

manual

[°C]

- 20 ... + 130

1