Resetting to default settings
27
ba75675e01
01/2007
Resetting to default settings
You can reset the measurement parameters and the configuration to the
delivery status separately from one another (initialization).
Measurement
parameters
The following measurement parameters can be reset to the delivery status:
Configuration
parameters
The following configuration parameters (InI) can be reset to the delivery
status:
Reset procedure:
Measuring mode
pH
Slope
-59.16 mV/pH
Asymmetry
0 mV
Manual temperature input
25 °C
Baud rate
4800
Display during pH calibration
Buffer nominal value
Interval 1
(automatically saved)
OFF
Interval 2
(for data transmission)
OFF
z
Press <CAL> while pressing the <RUN/ENTER> key.
z
To reset the measurement parameters, use <
▲
> <
▼
> to select YES and
confirm with <RUN/ENTER>, or
z
Continue to the configuration parameters without resetting (InI) with the
<RUN/ENTER> key.
z
After the InI configuration parameters, the instrument switches to the last
active measurement mode.