Xylem PH1970i PORTABLE PH METER User Manual
Page 23
Advertising
pH 1970i
Calibration
23
ba75675e01
01/2007
z
Immerse the pH electrode in the second buffer solution to continue the
two-point calibration.
z
When measuring without a temperature sensor:
Enter the temperature of the third buffer solution with <
▲
> <
▼
>.
z
Press <RUN/ENTER>. The AutoRead measurement begins. The
electrode voltage (mV) or the buffer nominal value appears on the
display.
z
When the measured value is stable, the instrument shows the value of
the slope and the evaluation of the three-point calibration.
z
Press <RUN/ENTER>. The instrument displays the value of the
asymmetry.
z
Switch to the measuring mode with <M>.
-