pH 1970i

Calibration

23

ba75675e01

01/2007

z

Immerse the pH electrode in the second buffer solution to continue the
two-point calibration.

z

When measuring without a temperature sensor:
Enter the temperature of the third buffer solution with <

> <

>.

z

Press <RUN/ENTER>. The AutoRead measurement begins. The
electrode voltage (mV) or the buffer nominal value appears on the
display.

z

When the measured value is stable, the instrument shows the value of
the slope and the evaluation of the three-point calibration.

z

Press <RUN/ENTER>. The instrument displays the value of the
asymmetry.

z

Switch to the measuring mode with <M>.

-

m

cm

S/

%

mg/ l

Sal

O

S

pH

5 8 8

pH

mV/

1

pH

cm

S/

mg/ l

Sal

O

S

pH

2

mV