6 ph value, 1 general information, Ph value 4.6.1 – Xylem Multi 3410 User Manual
Page 38: General information

Operation
Multi 3410
38
ba75863e11
02/2011
4.6
pH value
4.6.1
General information
You can measure the following parameters:
z pH value [ ]
z Sensor voltage [mV]
Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This enables interference-free measurements in
grounded measuring media.
Temperature
measurement
For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample. IDS sensors measure the temperature with
a temperature sensor integrated in the IDS sensor.
When operating a sensor without integrated temperature sensor, e.g.
via an IDS pH adapter, you have to determine and input the tempera-
ture of the measuring medium first.
Preparatory activities
Perform the following preparatory activities when you want to measure:
1
Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.
2
If necessary, select the measured parameter with <M>.
3
If necessary, calibrate or check the IDS pH sensor.