6 ph value, 1 general information, Ph value 4.6.1 – Xylem Multi 3410 User Manual

Page 38: General information

Advertising
background image

Operation

Multi 3410

38

ba75863e11

02/2011

4.6

pH value

4.6.1

General information

You can measure the following parameters:

z pH value [ ]

z Sensor voltage [mV]

Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This enables interference-free measurements in
grounded measuring media.

Temperature

measurement

For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample. IDS sensors measure the temperature with
a temperature sensor integrated in the IDS sensor.

When operating a sensor without integrated temperature sensor, e.g.
via an IDS pH adapter, you have to determine and input the tempera-
ture of the measuring medium first.

Preparatory activities

Perform the following preparatory activities when you want to measure:

1

Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.

2

If necessary, select the measured parameter with <M>.

3

If necessary, calibrate or check the IDS pH sensor.