Xylem Multi 3410 User Manual
Page 71

Multi 3410
Operation
71
ba75863e11
02/2011
FDO Check /
Check interval
1 ... 60 ...
999 d
Interval for the FDO Check (in days).
The FDO Check status display indica-
tor in the measuring window reminds
you to check the sensor regularly.
Sal correction
on
off
Manual salt content correction for con-
centration measurements
Salinity
0.0 ...
70.0
Salinity or salinity equivalent for the
salt content correction.
This function is only available for con-
centration measurements if the man-
ual salt content correction is switched
on.
Response time
t90
30 ... 300
Response time of the signal filter (in
seconds).
A signal filter in the sensor reduces the
fluctuation range of the measured
value. The signal filter is characterized
by the response time t90. This is the
time after which 90 % of a signal
change is displayed.
This menu item is only available if sen-
sor and meter support this function.
Firmware updates for the IDS sensor
and the meter are provided on the
internet (see chapter 9).
Reset
-
Resets all sensor settings to the deliv-
ery condition (see section 4.12.1)
Menu item
Possible
setting
Description